Key result
The Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices failed to meet the International Protocol criteria for accuracy compared to a mercury sphygmomanometer.
Why the study?
Do the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
Observational (n=54)
Do the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
The Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices do not meet International Protocol accuracy criteria and are not recommended for clinical use in adults.
Answers are grounded in this paper and cite their sources.
Avoid these wrist devices clinically; leaves open accuracy of other unvalidated monitors in routine practice.
OBJECTIVE: A variety of automatic blood measurement devices with diverse features have been introduced to the medical markets recently. Among these devices, models that measure at the wrist have become increasingly popular in self measurements. The objective of this study was to evaluate the accuracy of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices against the mercury sphygmomanometer in adults according to the International Protocol criteria. METHODS: Fifty-four patients over 30 years of age were studied and classified based on the International Protocol range. Blood pressure measurements at the wrist with the Samsung SBM-100A and Microlife BP 3BU1-5 were compared with the results obtained by two trained observers using a mercury sphygmomanometer. Nine sequential blood pressure measurements were taken. A total of 33 participants with randomly distributed arm circumferences were selected for both of the validation studies. During each validation study, 99 measurements were obtained for comparison from 33 participants. The first phase was performed on 15 participants and if the device passed this phase, 18 more participants were selected. RESULTS: Mean discrepancies and standard deviations of the device-sphygmomanometer were 0.9+/-9.2 and -2.7+/-9.3 mmHg for systolic blood pressure and -1.4+/-8.0 mmHg and 1.4+/-5.7 for diastolic blood pressure in the Samsung and Microlife study groups, respectively. The Samsung SBM-100A passed Phase 1 in 15 participants. Despite the fact that Microlife BP 3BU1-5 passed Phase 1 for diastolic pressure, it failed according to the systolic pressure criteria. Eighteen patients were added and Phase 2 was continued, in which Samsung SBM-100A failed to meet the criteria of Phases 2.1 and 2.2 for adults in systolic and diastolic blood pressure. It was found that the Microlife BP 3BU1-5 does not meet the criteria of either of Phases 2.1 and 2.2 for systolic blood pressure and Phase 2.2 for diastolic blood pressure. CONCLUSION: In this study, Samsung SBM-100A and Microlife blood pressure 3BU1-5 wrist blood pressure monitoring devices were found to be incompetent to meet the criteria of the International Protocol and it has not been possible to suggest any one of them for clinical use in adults.
No takes yet. Share an insight, caveat, or question.
Altunkan et al. (2007) conducted an observational in Blood pressure measurement (n=54). Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices vs. Mercury sphygmomanometer was evaluated on Device accuracy according to the International Protocol criteria. The Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices failed to meet the International Protocol criteria for accuracy compared to a mercury sphygmomanometer.