Key result
The Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices failed to meet the International Protocol criteria for accuracy compared to a mercury sphygmomanometer.
Why the study?
Do the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
Observational (n=54)
Do the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
The Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices do not meet International Protocol accuracy criteria and are not recommended for clinical use in adults.
No takes yet. Share an insight, caveat, or question.
Avoid these wrist devices clinically; leaves open accuracy of other unvalidated monitors in routine practice.
Altunkan et al. (2007) conducted an observational in Blood pressure measurement (n=54). Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices vs. Mercury sphygmomanometer was evaluated on Device accuracy according to the International Protocol criteria. The Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure devices failed to meet the International Protocol criteria for accuracy compared to a mercury sphygmomanometer.