Key result
The Microlife BP W200-1 wrist device accurately measured blood pressure, with mean differences between device and observer of -1.2+/-4.5 mmHg for SBP and -0.2+/-3.1 mmHg for DBP.
Why the study?
Does the Microlife BP W200-1 wrist device accurately measure blood pressure compared to observer measurement in adults?
Observational
Does the Microlife BP W200-1 wrist device accurately measure blood pressure compared to observer measurement in adults?
The Microlife BP W200-1 wrist monitor satisfies the European Society of Hypertension accuracy levels and can be recommended for clinical use in adults.
Answers are grounded in this paper and cite their sources.
Supports device validation for adult use; leaves open need for outcome trials in broader settings.
OBJECTIVE: The objective of the study was to determine the accuracy of the BP W200-1 device for blood pressure measurement at the wrist developed by the Microlife Company. The BP W200-1 model features a 'Microlife Average Mode', which uses a 'weighted' average of a minimum of three consecutive blood pressure (BP) readings taken 15 s apart. METHOD: Device evaluation was carried out using the International Protocol of the European Society of Hypertension. Monitor performance was assessed in relation to patients' sex, age, wrist circumference, and systolic (SBP) and diastolic blood pressures (DBP). RESULTS: The BP W200-1 comfortably passed all three phases of the protocol for SBP and DBP. The device performed well in all ranges of SBP and DBP. Mean blood pressure differences for the BP W200-1 monitor were -1.2+/-4.5 mmHg for SBP and -0.2+/-3.1 mmHg for DBP. In multivariable analyses, the SBP and DBP discrepancies between device and observer were unrelated to age, sex, wrist circumference, and blood pressure level. CONCLUSION: These data show that the Microlife BP W200-1 wrist monitor satisfies the recommended European Society of Hypertension accuracy levels for SBP and DBP and therefore can be recommended for clinical use in the adult population. Its performance is uniform across subgroups of patients with different clinical characteristics.
No takes yet. Share an insight, caveat, or question.
Palatini et al. (2008) conducted an observational in Blood pressure measurement. Microlife BP W200-1 wrist device vs. Observer was evaluated on Accuracy of blood pressure measurement according to the International Protocol of the European Society of Hypertension (Mean difference: SBP -1.2 mmHg, DBP -0.2 mmHg). The Microlife BP W200-1 wrist device accurately measured blood pressure, with mean differences between device and observer of -1.2+/-4.5 mmHg for SBP and -0.2+/-3.1 mmHg for DBP.
Synapse has enriched 10 related papers. Consider them for comparative context: