Why the study?
Does the Masimo SET Rad-9 pulse oximeter improve photoplethysmographic signal phase characteristics compared to the Novametrix Oxypleth in healthy adults?
Population
9 healthy adults
Comparison
Masimo SET Rad-9 pulse oximeter vs Novametrix Oxypleth pulse oximeter
Design
Other, PPG signals were acquired from fingers on the same hand in a random…
Authors
Loading...
Masimo oximeter's greater PTT variability cautions against interchangeable use for timing metrics; leaves open optimal device selection for photoplethysmographic cardiovascular monitoring.
Does the Masimo SET Rad-9 pulse oximeter improve photoplethysmographic signal phase characteristics compared to the Novametrix Oxypleth in healthy adults?
Raw photoplethysmography signals obtained from different modern pulse oximeters exhibit significant variability in phase delay, which can substantially affect the accuracy of timing-related cardiovascular measurements like pulse transit time.
Foo et al. (2005) studied this question.