Why the study?
Are low-cost oscillometric non-invasive blood pressure monitors repeatable and consistent between devices?
Are low-cost oscillometric non-invasive blood pressure monitors repeatable and consistent between devices?
Low-cost oscillometric blood pressure monitors demonstrate good within-device repeatability but exhibit significant between-device differences that could be misinterpreted clinically.
Answers are grounded in this paper and cite their sources.
Inter-device variability may affect clinical BP interpretation; leaves open standardization needs for low-cost monitors in humans.
Automated non-invasive blood pressure measuring devices based on the oscillometric technique are used widely for self-measurement and are often used in clinics in place of the manual, auscultatory method. Oscillometry was originally developed for monitoring purposes and there are questions over its suitability for making diagnostic measurements. This study measured the differences between automated devices, in the absence of physiological variability. We studied 19 low-cost, automated, non-invasive blood pressure devices, using a repeatable artificial arm simulator, and measured the within-device repeatability and between-device differences. We found that the devices were repeatable (mean within-device difference 1 mmHg), but between-device differences were 4.4 mmHg (systolic pressure) and 3.6 mmHg (diastolic pressure), for normal and high-normal blood pressures. Individual devices are sufficiently repeatable for clinical trend use, but differences between devices are sufficiently large that they may be misinterpreted as clinically significant.
No takes yet. Share an insight, caveat, or question.
Sims et al. (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: