Key result
A baseline QRS duration greater than 160 ms was a significant predictor of requiring an optimized interventricular delay value different from simultaneous pacing.
Cohort (n=77)
In patients receiving cardiac resynchronization therapy, a baseline QRS duration >160 ms strongly predicts the need for echocardiography-guided optimization of interventricular delay.
Answers are grounded in this paper and cite their sources.
QRS >160 ms may flag CRT patients for non-simultaneous VV optimization; leaves open outcome benefit in prospective trials.
AIM: Cardiac resynchronization is a well tolerated and effective therapy for heart failure, but 30% of patients still do not respond to biventricular pacing. Optimization of device settings, in particular interventricular delay value, represents a plausible target for improving these results, but available literature is discordant. We aimed our study at the identification of the best suitable candidates to interventricular delay optimization. METHODS: A total of 77 consecutive patients with optimized drugs therapy underwent clinical, echocardiographic and electrocardiographic evaluation before and after 6 months from implantation of a biventricular defibrillator in accordance to current guidelines. In each patient, atrioventricular and interventricular delay values were optimized at predischarge with echocardiogram. RESULTS: The only predictor of an optimized interventricular delay value different from simultaneous (i.e. standard shipment setting), at both univariate and multivariate analyses, was a QRS duration greater than 160 ms (odds ratio 22.958; P = 0.003) with a sensitivity of 70.9%. CONCLUSION: Candidates to cardiac resynchronization therapy with a basal QRS greater than 160 ms have a higher chance of requiring echo-guided tailoring of interventricular delay value. A strategy based on these data can potentially improve device programming, reducing by one-third the need for optimization, according to our findings, and at the same time avoid unnecessary time-consuming procedures.
No takes yet. Share an insight, caveat, or question.
Ziacchi et al. (2014) conducted a cohort in Heart failure (n=77). QRS duration > 160 ms was evaluated on Optimized interventricular delay value different from simultaneous (OR 22.958, p=0.003). A baseline QRS duration greater than 160 ms was a significant predictor of requiring an optimized interventricular delay value different from simultaneous pacing.
Synapse has enriched 10 related papers. Consider them for comparative context: