Why the study?
Does a linear logistic model incorporating clinical and signal-averaged electrocardiographic variables improve the prediction of ventricular tachycardia during electrophysiologic testing compared to established late-potential criteria in patients undergoing EP testing?
Does a linear logistic model incorporating clinical and signal-averaged electrocardiographic variables improve the prediction of ventricular tachycardia during electrophysiologic testing compared to established late-potential criteria in patients undergoing EP testing?
Answers are grounded in this paper and cite their sources.
A linear logistic model combining clinical and signal-averaged ECG variables improves the sensitivity for predicting ventricular tachycardia during electrophysiologic testing compared to standard late-potential criteria.
To improve the predictive accuracy of the signal-averaged electrocardiogram, we created a linear logistic model for predicting ventricular tachycardia during electrophysiologic testing. This signal-averaged electrocardiographic model was created from data obtained from 214 patients undergoing electrophysiologic testing (70 had ventricular tachycardia during electrophysiologic testing) by using stepwise logistic regression to rank eight clinical and nine signal-averaged electrocardiographic variables. The best predictors were ejection fraction, history of infarction, ventricular ectopic pairs or nonsustained ventricular tachycardia on Holter monitoring, QRS duration after 25-Hz filtering, and root mean square voltage of the terminal 40 msec of the QRS complex after 40- and 80-Hz filtering. Cross validation (a statistical technique that can be used to accurately evaluate how a predictive model will perform on a prospective patient population) was used to validate the model. After cross validation, the model's sensitivity was 91% and specificity was 59% for predicting ventricular tachycardia during electrophysiologic testing. This model compared favorably with established 25-Hz late-potential criteria (QRS duration of more than 110 msec and root mean square voltage of less than 25 microV of the terminal 40 msec of the QRS complex; sensitivity, 64%; specificity, 85%) and with established 40-Hz late-potential criteria (QRS duration of more than 114 msec or root mean square voltage of less than 20 microV of the terminal 40 msec of the QRS complex or duration of the low-amplitude signal less than 40 microV at the terminal QRS complex that is greater than 38 msec; sensitivity, 84%; specificity, 54%).(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Vatterott et al. (1990) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: