Answers are grounded in this paper and cite their sources.
In this paper, we present an ultrahighly sensitive methane (CH 4 ) sensor based on tunable diode laser absorption spectroscopy (TDLAS), utilizing an 80-m optical path length (OPL) multipass cell (MPC) with 15 independent circles pattern. The designed MPC achieved an actual OPL of 81.58 m within a compact volume of 361 cm 3, resulting in an exceptional ratio of optical path length to volume (RLV) of 22.6 cm –2 . Wavelength modulation spectroscopy (WMS) was employed to minimize background noise and enhance detection performance. The CH 4 -TDLAS sensor demonstrated excellent linear response to CH 4 concentrations. The minimum detection limit (MDL) of the sensor was determined to be 13.46 ppb. Further analysis using Allan deviation revealed that the MDL improved to 1.36 ppb at an average time of 400 s, highlighting the sensor’s exceptional long-term stability and ultrahigh sensitivity. Finally, to further validate the sensor’s performance, an experiment simulating laboratory CH 4 leakage was conducted. The results demonstrated that the CH 4 -TDLAS sensor exhibits ultrahigh sensitivity, effectively detecting and tracking changes in CH 4 concentration in real time.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: