Why the study?
Data including a detailed analysis of transcatheter heart valve positioning after TAVI in bicuspid aortic valves were limited.
Does TAVI with the balloon-expandable Sapien 3 prosthesis achieve similar prosthesis positioning in bicuspid aortic valves compared to tricuspid aortic valves?
Does TAVI with the balloon-expandable Sapien 3 prosthesis achieve similar prosthesis positioning in bicuspid aortic valves compared to tricuspid aortic valves?
TAVI with the Sapien 3 prosthesis in bicuspid aortic valves achieves accurate positioning comparable to tricuspid valves, although with slightly lower expansion and more pronounced waist.
Answers are grounded in this paper and cite their sources.
May support similar SAPIEN 3 positioning targets in BAV; leaves open whether greater waist and underexpansion affect durability.
Background: Prior data suggest a correlation between the position of transcatheter heart valves (THV) and the occurrence of complications after transcatheter aortic valve implantation (TAVI) in patients with tricuspid aortic valves (TAV). However, data including a detailed analysis of prosthesis positioning in bicuspid aortic valves (BAV) are limited. Therefore, the purpose of this study was to investigate THV position after TAVI in BAV. Methods: We evaluated the THV position in 50 BAV and 50 TAV patients (all received the balloon-expandable Sapien 3 prosthesis) using fusion imaging of pre- and post-procedural computed tomography angiography. According to the manufacturers’ recommendations, a low implantation position was defined as >30% of the prosthesis below the annulus. Results: THV position was appropriate in the majority of the patients within both groups (90.0% for BAV vs. 96.0% for TAV, p = 0.240). In BAV, we observed a more pronounced THV waist (7.4 ± 4.5% vs. 5.8 ± 3.0%, p = 0.043) and a lower average THV expansion (91.9 ± 12.2% vs. 95.5 ± 2.7% of nominal expansion, p = 0.044). Conclusions: Accurate positioning in relation to the aortic annulus of the TAVI Sapien 3 prosthesis is possible in patients with BAV with results comparable to TAV. However, there is a more pronounced prosthesis waist and a lower average THV expansion in BAV.
No takes yet. Share an insight, caveat, or question.
Breitbart et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: