Why the study?
Active-fixation leads are associated with higher rates of cardiac perforation, but large series evaluating radiographic predictors of right ventricular lead perforation are lacking.
What are the fluoroscopic predictors of right ventricular perforation caused by active-fixation pacing and defibrillator leads?
What are the fluoroscopic predictors of right ventricular perforation caused by active-fixation pacing and defibrillator leads?
Fluoroscopic parameters such as lead slack score and changes in lead slack height and width, along with LVEF, are independent predictors of right ventricular perforation from active-fixation leads.
Answers are grounded in this paper and cite their sources.
Fluoroscopic slack metrics may flag higher RV perforation risk during implantation; leaves open prospective validation before guiding technique.
INTRODUCTION: Active-fixation leads have been associated with higher incidence of cardiac perforation. Large series specifically evaluating radiographic predictors of right ventricular (RV) lead perforation are lacking. METHODS: We conducted a retrospective observational study including 1691 consecutive patients implanted with an active fixation pacing and defibrillator lead at our institution between January 2015 and January 2021. Fourteen patients who had clinically relevant RV perforation caused by pacemaker and implantable cardioverter-defibrillator leads were included in the study. RESULTS: Univariate and multivariate analyses were used to identify predictors of RV perforation. In multivariate analysis, lead slack score (odds ratio [OR]: 3.694, 95% confidence interval [CI]: 1.066-12.807; p = .039), change in lead slack height (OR: 1.218, 95% CI: 1.011-1.467; p = .038) and width (OR: 1.253, 95% CI: 1.120-1.402; p = .001), left ventricular ejection fraction (OR: 0.995, 95% CI: 0.910-1.088; p = .032) were independent predictors of RV perforation. CONCLUSION: Fluoroscopic predictors of RV perforation associated with RV lead can be easily determined during implantation. Identification of these predictors may prevent the sequelae of RV perforation associated with active-fixation leads.
No takes yet. Share an insight, caveat, or question.
Çelikyurt et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: