Key result
Heated saline needle ablation achieves ~97% acute clinical VT non-inducibility.
Why the study?
Ventricular tachycardia ablation is limited by the inability to create penetrating lesions to reach intramyocardial origins, prompting evaluation of intramural needle ablation using heated saline-enhanced radiofrequency energy.
Does intramyocardial needle-delivered heated saline-enhanced radiofrequency (SERF) ablation safely eliminate clinical ventricular tachycardia inducibility and reduce VT burden in patients with drug- and ablation-refractory VT?
Does intramyocardial needle-delivered heated saline-enhanced radiofrequency (SERF) ablation safely eliminate clinical ventricular tachycardia inducibility and reduce VT burden in patients with drug- and ablation-refractory VT?
First-in-human feasibility data demonstrate that intramyocardial heated saline-enhanced radiofrequency needle ablation achieves 97% acute clinical VT non-inducibility and an 89% reduction in ICD therapies in patients with refractory VT who failed prior conventional ablations, though notable periprocedural thromboembolic and hemodynamic risks require careful procedural titration.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for SERF ablation in refractory VT; randomized trials needed before clinical adoption.
BACKGROUND: Ablation of ventricular tachycardia (VT) is limited by the inability to create penetrating lesions to reach intramyocardial origins. Intramural needle ablation using in-catheter, heated saline-enhanced radio frequency (SERF) energy uses convective heating to increase heat transfer and produce deeper, controllable lesions at intramural targets. This first-in-human trial was designed to evaluate the safety and efficacy of SERF needle ablation in patients with refractory VT. METHODS: Thirty-two subjects from 6 centers underwent needle electrode ablation. Each had recurrent drug-refractory monomorphic VT after implantable cardioverter defibrillator implantation and prior standard ablation. During the SERF study procedure, one or more VTs were induced and mapped. The SERF needle catheter was used to create intramural lesions at targeted VT site(s). Acute procedural success was defined as noninducibility of the clinical VT after the procedure. Patients underwent follow-up at 30 days, and 3 and 6 months, with implantable cardioverter defibrillator interrogation at follow-up to determine VT recurrence. RESULTS: These refractory VT patients (91% male, 66±10 years, ejection fraction 35±11%; 56% ischemic, and 44% nonischemic) had a median of 45 device therapies (shock/antitachycardia pacing) for VT in the 3 to 6 months pre-SERF ablation. The study catheter was used to deliver an average of 10±5 lesions per case, with an average of 430±295 seconds of radiofrequency time, 122±65 minute of catheter use time, and a procedural duration of 4.3±1.3 hours. Acute procedural success was 97% for eliminating the clinical VT. At average follow-up of 5 months (n=32), device therapies were reduced by 89%. Complications included 2 periprocedural deaths: an embolic mesenteric infarct and cardiogenic shock, 2 mild strokes, and a pericardial effusion treated with pericardiocentesis (n=1). CONCLUSIONS: Intramural heated saline needle ablation showed complete acute and satisfactory mid-term control of difficult VTs failing 1 to 5 prior ablations and drug therapy. Further study is warranted to define safety and longer-term efficacy. REGISTRATION: URL: https://www. CLINICALTRIALS: gov; Unique Identifier: NCT03628534 and NCT02994446.
No takes yet. Share an insight, caveat, or question.
Packer et al. (2022) studied Refractory monomorphic ventricular tachycardia due to structural heart disease (n=32). Intramyocardial needle-delivered heated saline-enhanced radiofrequency (SERF) ablation (Durablate Catheter and Thermedical Ablation System) was evaluated on Acute procedural success (non-inducibility of the targeted clinical VT at the end of the ablation procedure). Intramyocardial heated saline-enhanced radiofrequency needle ablation achieved 97% acute non-inducibility of clinical VT and reduced ICD device therapies by 89% at 5 months.
Synapse has enriched 10 related papers. Consider them for comparative context: