Key result
The ValveClamp transcatheter device reduced median mitral regurgitant maximum jet area to left atrial area from 57% to 0% in dogs with myxomatous mitral valve disease.
Why the study?
Veterinary management of canine mitral regurgitation from myxomatous disease is limited to medical therapy, and human surgical options require cardiopulmonary bypass with high complication risks.
Does the ValveClamp edge-to-edge transcatheter device reduce mitral regurgitation severity in dogs with naturally occurring myxomatous mitral valve disease?
Observational (n=8)
No
Does the ValveClamp edge-to-edge transcatheter device reduce mitral regurgitation severity in dogs with naturally occurring myxomatous mitral valve disease?
Answers are grounded in this paper and cite their sources.
The ValveClamp edge-to-edge closure device safely and effectively reduces mitral regurgitation in a canine model of naturally occurring myxomatous mitral valve disease.
Background: Veterinary management of mitral valve regurgitation due to mxyomatous valve disease in dogs is limited to medical treatments, which only postpones the onset of congestive heart failure or alleviates clinical symptoms. Most surgical procedures to manage this condition in humans require cardiopulmonary bypass and have a high risk of complications. Animals: Eight dogs with naturally occurring mitral valve regurgitation. Methods: Prospective observational study. All dogs were treated with a novel edge-to-edge transcatheter device named ValveClamp. The total surgical procedural time and total catheterization time were recorded. Echocardiographic variables measured pre- and post-procedure were compared using Wilcoxin-signed rank test with a P < 0.05 considered significant. Data were expressed as median and interquartile range and absolute numbers and percentages. Results: The procedural success rate was 100% and all the dogs survived without complications. The median (interquartile range) total surgical procedural time was 86.5 (76–96.2) minutes and catheterization time was 23.5 (22–33.8) minutes. Echocardiography revealed a significant reduction in mitral regurgitation severity in all dogs following the procedure based on both a reduced mitral regurgitant maximum jet area ( P = 0.012) and a reduced mitral regurgitant maximum jet area to left atrial area ( P = 0.018). Conclusion: The ValveClamp device is effective at reducing the severity of mitral regurgitation in dogs with naturally occurring myxomatous valve disease.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2020) conducted an observational in Myxomatous mitral valve disease (MMVD) (n=8). ValveClamp edge-to-edge transcatheter device vs. Pre-procedure baseline was evaluated on Mitral regurgitant maximum jet area to left atrial area (MRA/LAA) (p=0.018). The ValveClamp transcatheter device reduced median mitral regurgitant maximum jet area to left atrial area from 57% to 0% in dogs with myxomatous mitral valve disease.
Synapse has enriched 10 related papers. Consider them for comparative context: