Key result
A heart rate variability-based seizure detection algorithm achieved 87.0% sensitivity and a false alarm rate of 0.9/24 hours in patients with marked ictal autonomic changes.
Why the study?
Although several validated seizure detection algorithms exist for convulsive seizures, detection of nonconvulsive seizures remains challenging.
Does an HRV-based seizure detection algorithm accurately detect seizures compared to video-EEG in patients with marked ictal autonomic changes?
Observational (n=19)
Does an HRV-based seizure detection algorithm accurately detect seizures compared to video-EEG in patients with marked ictal autonomic changes?
HRV-based seizure detection using a wearable ECG device shows high sensitivity and low false alarm rates in patients with marked ictal autonomic changes.
Answers are grounded in this paper and cite their sources.
Supports wearable HRV seizure detection in selected patients; leaves open prospective validation in broader epilepsy cohorts.
Although several validated seizure detection algorithms are available for convulsive seizures, detection of nonconvulsive seizures remains challenging. In this phase 2 study, we have validated a predefined seizure detection algorithm based on heart rate variability (HRV) using patient-specific cutoff values. The validation data set was independent from the previously published data set. Electrocardiography (ECG) was recorded using a wearable device (ePatch) in prospectively recruited patients. The diagnostic gold standard was inferred from video-EEG monitoring. Because HRV-based seizure detection is suitable only for patients with marked ictal autonomic changes, we defined responders as the patients who had a>50 beats/min ictal change in heart rate. Eleven of the 19 included patients with seizures (57.9%) fulfilled this criterion. In this group, the algorithm detected 20 of the 23 seizures (sensitivity: 87.0%). The algorithm detected all but one of the 10 recorded convulsive seizures and all of the 8 focal impaired awareness seizures, and it missed 2 of the 4 focal aware seizures. The median sensitivity per patient was 100% (in nine patients all seizures were detected). The false alarm rate was 0.9/24 h (0.22/night). Our results suggest that HRV-based seizure detection has high performance in patients with marked autonomic changes.
No takes yet. Share an insight, caveat, or question.
Jeppesen et al. (2020) conducted an observational in Seizures (n=19). Heart rate variability-based seizure detection algorithm vs. Video-EEG monitoring was evaluated on Seizure detection sensitivity in patients with >50 beats/min ictal change in heart rate. A heart rate variability-based seizure detection algorithm achieved 87.0% sensitivity and a false alarm rate of 0.9/24 hours in patients with marked ictal autonomic changes.