Can an opposed solenoid intravascular receiver coil generate high-quality in vivo MR images of the aorta in a porcine model?
Can an opposed solenoid intravascular receiver coil generate high-quality in vivo MR images of the aorta in a porcine model?
Intravascular MR imaging with restricted coil motion and blood signal saturation can produce high-quality in vivo images of the aortic wall in a porcine model.
Answers are grounded in this paper and cite their sources.
Supports feasibility in porcine aortic imaging; leaves open human translation and clinical validation.
An opposed solenoid intravascular receiver coil is used in a series of five pigs to generate images of the aorta. Severe ghosting artifacts are shown to occur if the motion of the intravascular coil is not restricted with respect to the vessel wall. Restricting coil motion and saturating the blood signal superior to the imaging plane are shown to produce high quality MR images. Layers within the artery wall could be discerned and the results of the in vivo analysis compared favorably with high resolution MR images of the corresponding tissue after excision.