Key result
USCOM stroke volume measurements demonstrated excellent inter-rater reliability between a newly trained and experienced operator on acute emergency in-patients.
Why the study?
Does the non-invasive ultrasonic cardiac output monitor (USCOM) demonstrate a short learning curve and high inter-rater reliability for stroke volume measurement in acute emergency admissions?
Observational (n=124)
Single-blind
No
Does the non-invasive ultrasonic cardiac output monitor (USCOM) demonstrate a short learning curve and high inter-rater reliability for stroke volume measurement in acute emergency admissions?
Following a training period of less than 5 hours, USCOM stroke volume measurements demonstrated excellent inter-rater reliability for assessing fluid responsiveness in acute emergency admissions.
Answers are grounded in this paper and cite their sources.
Supports feasibility of brief USCOM training for reliable measurements in emergencies; leaves open clinical utility for guiding fluid therapy.
Traditionally, assessment of the cardiac output has been limited to theatre or the intensive care unit. However, non-invasive cardiac output estimation is now readily available, and its application may have wider benefit in the emergency setting. The non-invasive ultrasonic cardiac output monitor (USCOM) was investigated to determine its learning curve and inter-rater reliability. Four trainee operators each performed stroke volume measurements on 25 volunteers, compared to an experienced operator pre- and post-passive leg raise. Inter-rater reliability was then assessed on 24 acute emergency in-patients. Mean percentage difference in stroke volume decreased from 19% (95% confidence intervals 14–23) across volunteers 1–5, to 6% (4–8) for the last 5 volunteers scanned. Consequently, on acute emergency in-patients, excellent inter-rater reliability (Lin’s concordance correlation coefficient (ρ c ) 0.96 (0.92–0.98)) and agreement of a change ≥10% in stroke volume following passive leg raise on 23/24 cases were found. Following a training period of less than 5 h, USCOM stroke volume measurements demonstrated excellent inter-rater reliability.
No takes yet. Share an insight, caveat, or question.
Hodgson et al. (2015) conducted an observational in Acute emergency admissions (n=124). Non-invasive ultrasonic cardiac output monitor (USCOM) vs. Experienced operator was evaluated on Inter-rater reliability of stroke volume measurements (rc 0.96, 95% CI 0.92-0.98). USCOM stroke volume measurements demonstrated excellent inter-rater reliability between a newly trained and experienced operator on acute emergency in-patients.
Synapse has enriched 10 related papers. Consider them for comparative context: