Electrical leakage current from defective grounding in medical equipment can inhibit implanted demand pacemakers, highlighting a critical safety hazard.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to verify equipment grounding near pacemakers; leaves open systematic safety assessments in device patients.
A 76-year-old man had his implanted demand pacemaker inhibited by an electrical leakage current caused by defective grounding in an electrocardiographic recorder power cable. Various measures can be undertaken in order to minimize this hazard, and these are discussed.
No takes yet. Share an insight, caveat, or question.
Meibom et al. (1971) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: