Why the study?
Can a subcutaneous ICD be successfully implanted without device interference in a patient with an existing LVAD?
Can a subcutaneous ICD be successfully implanted without device interference in a patient with an existing LVAD?
Answers are grounded in this paper and cite their sources.
This case report provides the first evidence that a subcutaneous ICD can be successfully and safely implanted in a patient with an existing LVAD without device interference.
A 56-year-old man with ischemic cardiomyopathy, a biventricular implantable cardioverter-defibrillator (ICD), and a left ventricular assist device (LVAD) developed a pocket hematoma and infection after an ICD generator change. The biventricular ICD was extracted, and the patient was given a full course of antibiotics. Because he had no indications for bradycardia pacing or biventricular pacing, he was implanted with a subcutaneous ICD under full anticoagulation. There was no interference in sensing or shock delivery from the ICD. The LVAD readings were unchanged during and after the procedure. The patient had an uneventful postoperative course, and both devices were functioning normally. To our knowledge, this is the first reported case of the implantation of a subcutaneous ICD in the presence of an LVAD. This report illustrates that both devices can be implanted successfully in the same patient. In addition, the subcutaneous ICD minimizes the risk of bloodstream infections, which can be fatal in patients who have life-supporting devices such as an LVAD.
No takes yet. Share an insight, caveat, or question.
Gupta et al. (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: