Why the study?
The study was conducted to evaluate the accuracy of a wrist-based heart rate monitor at different exercise intensities across different skin tones.
Does wrist-based photoplethysmography accurately measure heart rate across different skin tones and exercise intensities compared to a chest strap?
Does wrist-based photoplethysmography accurately measure heart rate across different skin tones and exercise intensities compared to a chest strap?
Wrist-based heart rate monitors exhibit significantly greater measurement error at higher exercise intensities in individuals with darker skin tones.
May warrant caution interpreting wrist PPG HR at high intensity in darker skin tones; leaves open need for prospective validation before clinical adoption.
PURPOSE: To evaluate the accuracy of a wrist-based heart rate (HR) monitor at different exercise intensities across different skin tones. METHODS: Using a cross-sectional design, we compared HR measures from the wrist-based photoplethysmography Fitbit Charge 5 to the Polar H10 chest strap at rest and during the YMCA Protocol using a recumbent cycle ergometer. Participant were grouped into three skin tone categories: light (Fitzpatrick Scale Skin Types 1+2), medium (Types 3+4), and darker skin tone (Types 5+6). HR measures using the Polar chest strap during the exercise test were categorized as <40%, 40-60%, or >60% HR reserve (HRR). Absolute error in beats per minute (bpm) between the two devices was calculated for each measure. A linear mixed effects model was used to assess interaction effects between skin tone and exercise intensity, with participants as the random effect. Bland-Altman plots were used for visual analyses. RESULTS: Twenty-five participants [mean (SD): 25.8 (1.9) years old; 64% female] were included with 495 observations of simultaneous Fitbit and Polar HR recordings collected during exercise. During exercise, we observed a statistically significant interaction effect between skin tone and exercise intensity. Compared with light skin tone at <40% HRR, mean error was greater for medium skin tone at >60% HRR [mean error (95%CI): 11.8 (5.6-17.9) bpm, p<0.001] and darker skin tone at 40-60% HRR [7.6 (1.7-13.5) bpm, p = 0.011] and >60% HRR [11.7 (5.3-18.0) bpm, p<0.001]. CONCLUSION: HR measurement error using a wrist-based device was greater with increasing exercise intensity for people with darker skin tones.
No takes yet. Share an insight, caveat, or question.
Hung et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: