Why the study?
Does dual-chamber ICD detection enhancement reduce inappropriate detection of SVT compared to single-chamber detection in patients with an indication for an ICD?
Does dual-chamber ICD detection enhancement reduce inappropriate detection of SVT compared to single-chamber detection in patients with an indication for an ICD?
Dual-chamber ICD detection enhancements significantly reduce the inappropriate detection of supraventricular tachycardia compared to single-chamber detection.
Dual-chamber detection enhancements reduce inappropriate SVT detection in ICD patients; confirms benefit versus single-chamber in an RCT.
BACKGROUND: Delivery of inappropriate shocks caused by misdetection of supraventricular tachycardia (SVT) remains a substantial complication of implanted cardioverter/defibrillator (ICD) therapy. Whether use of optimally programmed dual-chamber ICDs lowers this risk compared with that in single-chamber ICDs is not clear. METHODS AND RESULTS: Subjects with a clinical indication for ICD (n=400) at 27 participating centers received dual-chamber ICDs and were randomly assigned to strictly defined optimal single- or dual-chamber detection in a single-blind manner. Programming minimized ventricular pacing. The primary end point was the proportion of SVT episodes inappropriately detected from the time of programming until crossover or end of study. On a per-episode basis, 42% of the episodes in the single-chamber arm and 69% of the episodes in the dual-chamber arm were due to SVT. Mortality (3.5% in both groups) and early study withdrawal (14% single-chamber, 11% dual-chamber) were similar in both groups. The rate of inappropriate detection of SVT was 39.5% in the single-chamber detection arm compared with 30.9% in the dual-chamber arm. The odds of inappropriate detection were decreased by almost half with the use of the dual-chamber detection enhancements (odds ratio, 0.53; 95% confidence interval, 0.30 to 0.94; P=0.03). CONCLUSIONS: Dual-chamber ICDs, programmed to optimize detection enhancements and to minimize ventricular pacing, significantly decrease inappropriate detection.
No takes yet. Share an insight, caveat, or question.
Friedman et al. (2006) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: