Key result
Single-chamber ICD cutoff >=200 bpm cuts inappropriate shocks by ~3 percentage points vs >=170 bpm.
Why the study?
ICD programming influences inappropriate shock rates, motivating this post-hoc analysis of appropriate and inappropriate shocks by detection zones in the OPTION trial.
RCT (n=453)
Single-blind
1:1, 4-block permutation randomization
Yes
Answers are grounded in this paper and cite their sources.
Dual-chamber discrimination algorithms may mitigate inappropriate shocks at lower detection cutoffs; leaves open whether PARAD+ warrants randomized comparison against standard high-rate programming.
The programming of implantable cardioverter-defibrillators (ICDs) influences inappropriate shock rates. The aim of the study is to analyse rates of patients with appropriate and inappropriate shocks according to detection zones in the OPTION trial. All patients received dual chamber (DC) ICDs randomly assigned to be programmed either to single chamber (SC) or to DC settings including PARAD+ algorithm. In a post-hoc analysis, rates of patients with inappropriate and appropriate shocks were calculated for shocks triggered at heart rates ≥ 170 bpm (ventricular tachycardia zone) and at rates ≥ 200 bpm (ventricular fibrillation zone). In the SC group, higher rates of patients with total and inappropriate shocks were delivered at heart rates ≥ 170 bpm than at rates ≥ 200 bpm (total shocks: 21.1% vs. 16.6%; p = 0.002; inappropriate shocks: 7.6% vs. 4.5%, p = 0.016; appropriate shocks: 15.2% vs. 13.5%; p = n.s.). No such differences were observed in the DC group (total shocks: 14.3% vs. 12.6%; p = n.s.; inappropriate shocks: 3.9% vs. 3.6%; p = n.s.; appropriate shocks: 12.2% vs. 10.4%; p = n.s.). The higher frequency of patients with total shocks with SC settings than with DC settings that benefit from PARAD+ was driven by a higher percentage of patients with inappropriate shocks in the VT zone (170-200 bpm) in the SC population.
No takes yet. Share an insight, caveat, or question.
Kolb et al. (2016) conducted an RCT in Primary or secondary prevention of sudden cardiac death (heart failure with LVEF <= 40%) (n=453). Single-chamber (SC) ICD settings vs Dual-chamber (DC) ICD settings with PARAD+ and SafeR algorithms analyzed across detection zones (>=170 bpm vs >=200 bpm) vs. Shocks triggered at heart rates >=200 bpm (VF zone) vs >=170 bpm (VT+VF zones) within SC and DC ICD programming groups was evaluated on Percentage of patients receiving inappropriate ICD shocks triggered at heart rates >=170 bpm versus >=200 bpm in the single-chamber (SC) group (p=0.016). Single-chamber ICD programming delivered inappropriate shocks to more patients at >=170 bpm than >=200 bpm, whereas dual-chamber programming with PARAD+ showed no difference.
Synapse has enriched 10 related papers. Consider them for comparative context: