Why the study?
Does an ultrasound transit time (USTT) flow probe improve the accuracy of blood flow measurement at the pump head of centrifugal pumps compared to US-Doppler and electromagnetic systems?
Population
Centrifugal blood pumps (in vitro and in vivo models)
Comparison
Specially adapted flow probe based on the… vs Ultrasound Doppler systems and electromagnetic…
Design
Preclinical
Authors
Loading...
May enable precise centrifugal pump flow monitoring in MCS; leaves open clinical adoption pending in vivo validation.
Does an ultrasound transit time (USTT) flow probe improve the accuracy of blood flow measurement at the pump head of centrifugal pumps compared to US-Doppler and electromagnetic systems?
A specially adapted ultrasound transit time flow probe provides accurate blood flow measurements directly at the centrifugal pump housing, overcoming the limitations of Doppler and electromagnetic systems caused by high vorticity.
Schima et al. (1997) studied this question.