Answers are grounded in this paper and cite their sources.
A significant improvement in unaided visual standards may be achieved by switching from a monofocal to a 3M diffractive bifocal implant; however, this results in a loss of clarity, particularly for small print, in some patients. In a prospective study of 55 bifocal eyes and 55 monofocal eyes, best-case analysis showed 81% of 52 eyes with bifocal implants could read J5 with their distance correction compared with 21% of the control eyes and 56% could read J2 with the distance correction compared with 2% of the control eyes; 75% of bifocal eyes could read both 20/40 and J3 without any aid compared with 33% of the control eyes. The 3.5 diopter focus difference in the bifocal lens produced commendable results, with 52% requiring no spectacle reading addition. However, the fact that 17% of bifocal eyes could not read J2 highlights certain drawbacks that are relevant to indications for use.
No takes yet. Share an insight, caveat, or question.
Piers Percival (1990) studied this question.