Why the study?
Does the presence of a stent link at a side branch ostium affect stent strut malapposition and side branch jailing ratio during re-POT sequence in a coronary bifurcation bench model?
Does the presence of a stent link at a side branch ostium affect stent strut malapposition and side branch jailing ratio during re-POT sequence in a coronary bifurcation bench model?
In a bench model of coronary bifurcation stenting, the presence of a stent link at the side branch ostium did not significantly affect strut malapposition or side branch jailing ratio after a re-POT sequence.
Bench model shows no link effect on malapposition or jailing after re-POT; leaves open clinical relevance in bifurcation PCI.
OBJECTIVES: The purpose of this study was to compare the stent deformation, obstruction of stent struts at a jailed side branch (SB) ostium, and stent strut malapposition after a repetitive proximal optimizing technique (re-POT) sequence between bifurcation lesions with and without stent links at SB ostia in ex vivo experimental setting. METHODS: A flexible urethane coronary bifurcation bench model was used, and crossover single stent implantation was performed from main branch (MB) with re-POT sequence. Under videoscope observation, presence (link group, n = 12) or absence (no-link group, n = 12) of stent link at distal semicircle of SB ostium was intentionally set, and rewiring was performed through distal cell in four different stent platforms. RESULTS: There were no significant differences in the rate of malapposed struts and in SB jailing ratio after the re-POT sequence between the link and no-link groups. SB jailing ratio increased significantly from 8.0% to 9.6% during the second POT procedure (P < 0.001). SB jailing ratio after the second POT procedure differed among stent platforms. CONCLUSIONS: The presence of a stent link at an SB ostium was not associated with a rate of malapposed struts and SB jailing ratio after the re-POT sequence. SB jailing ratio was significantly increased after second POT procedure, but was different among stent platforms. The Xience stent might minimize the change of SB jailing ratio and be suitable for coronary bifurcation stenting using re-POT sequence.
No takes yet. Share an insight, caveat, or question.
Kume et al. (2018) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: