A retardation device situated after the final collector slit of a tandem magnetic mass spectrometer has improved the abundance sensitivity measured on the low mass side of a Cs+ peak from 1 × 106 to 4 × 109. On the high mass side the abundance sensitivity of approximately 109 is not affected significantly by the retardation technique. The abundance sensitivity extrapolated to the uranium mass region is improved from 5 × 105 to 109 on the low mass side and is 108 on the high mass side.