Answers are grounded in this paper and cite their sources.
A propeller-shaped region of reduced full-energy-peak efficiency was observed on the face of a true coaxial Ge(Li) detector. A similar pattern on the etch-pit photograph suggests that the propeller blades correspond to the 1 1 1 symmetry axes of the crystal. Spectra obtained when a 122 keV γ-ray beam was directed at this degraded region, show a low energy tail which is not present when the beam is directed at normal regions. Current pulses corresponding to γ-rays that interact in the propeller region have shapes that show deficiencies in the electron component. Thus, degraded crystal regions that are revealed by etch-pit distributions can be correlated to local detector performance by direct observation of trapping rather than only by inference from the presence of tails in the energy spectra.
No takes yet. Share an insight, caveat, or question.
Strauss et al. (1974) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: