Why the study?
What are the long-term survival and structural deterioration rates of the Ionescu-Shiley Low-Profile bioprosthetic heart valve in adults?
What are the long-term survival and structural deterioration rates of the Ionescu-Shiley Low-Profile bioprosthetic heart valve in adults?
The ISLP bioprosthetic valve shows increasing structural deterioration after 7 years, requiring regular surveillance and consideration of long-term anticoagulation due to higher than expected thrombo-embolic events.
Answers are grounded in this paper and cite their sources.
Higher mitral deterioration rates support enhanced post-MVR surveillance; leaves open durability comparisons across bioprostheses.
Seven hundred and eight adults (age > or = 16 years) with isolated aortic (n = 433) or mitral (n = 275) Ionescu-Shiley Low-Profile (ISLP) pericardial valves were followed at 14 implanting centres in Canada, the United Kingdom, and the United States for a mean of 6.7 years, providing 4,729 patient-years of clinical data. The operative mortality rate was 3.0% for aortic valve replacement (AVR) and 5.5% for mitral valve replacement (MVR) (p = ns). Actuarial patient survival following AVR at 5 years was 81.6%, and 62.9% at 10 years; for MVR patients it was 78.1% at 5 years and 59.6% at 10 years. The ISLP valve appears to have durability comparable to other contemporary bioprosthetic valves. For aortic prostheses, the freedom from structural deterioration was 96.5% at 5 years and 73.7% at 10 years, and 89.7% at 5 years and 62.4% at 10 years for mitral prostheses. Structural deterioration was significantly more frequent following MVR than after AVR (p < 0.05). Structural deterioration was the principal cause for reoperation, but sudden deterioration precluding safe reoperation was not a dominant feature of this series. The ISLP valve appeared to engender more thrombo-embolic events than would be anticipated from earlier studies of pericardial bioprostheses, but was indistinguishable from other tissue valves in its incidence of other valve-related complications. We conclude that ISLP valves now implanted for 7 years or more are entering a phase of increasing structural deterioration, indicating the need for regular clinical and echocardiographic surveillance, and that long-term anticoagulation should be instituted for relatively minimal indications in these patients.
No takes yet. Share an insight, caveat, or question.
Wheatley et al. (1994) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: