Answers are grounded in this paper and cite their sources.
Reveals that a mobile ECG device underestimates QTc intervals in drug-resistant mycobacterial infections, indicating its potential use as a preliminary screening tool before standard 12-lead ECG.
Multidrug resistant tuberculosis and non-tuberculous mycobacterium infections present challenges due to complex treatment regimens. Extended treatment regimes expose patients to higher risks of toxic side-effects. A high drug toxicity profile necessitates closer monitoring. One of the more challenging issues is QTc prolongation with non-injectable regimens. This study investigates the portable AliveCor device to record and measure the QTc on a 6-lead ECG. An automated QTc readout from 12-Lead ECG for each patient (n = 13) and mean QTc value calculated from each patients' respective AliveCor tracing were compared. The general trend suggests AliveCor underestimates QTc - 92% cases calculated the AliveCor QTc as lower than their corresponding 12-Lead QTc readout. The use of AliveCor could potentially be translated into current clinical practice with caution of percentage variation either side. This could facilitate the use of AliveCor as a promising and convenient screening tool before further evaluation by a 12-Lead ECG is required.
No takes yet. Share an insight, caveat, or question.
Puranik et al. (2021) studied this question.