PURPOSE: To assess the safety to the corneal endothelium of ultraviolet (UV) light treatment to a light-adjustable intraocular lens (IOL). SETTING: Codet Vision Institute, Tijuana, Mexico. DESIGN: Cohort study. METHODS: The study evaluated consecutive patients scheduled for phacoemulsification cataract surgery with implantation of a light-adjustable IOL. Endothelial cell density (ECD), variation in the size of endothelial cells, and the percentage of hexagonal cells were determined centrally with a specular microscope preoperatively, 1 week postoperatively (before UV treatment), and 6 months postoperatively. Two adjustments and 2 lock-in procedures were performed with a light-delivery system between 2 weeks and 4 weeks postoperatively. RESULTS: The study enrolled 10 patients. Specular microscopy showed a mean ECD of 2344 cells/mm(2) preoperatively. The mean overall central ECD was 2047 cells/mm(2) 1 week postoperatively and 2130 cells/mm(2) at 6 months, representing an endothelial cell loss of 12.6% and 9.1%, respectively, from preoperative values. There was no statistically significant difference in the coefficient of variation or percentage of hexagonal cells at either postoperative evaluation. CONCLUSIONS: The endothelial cell loss at 6 months correlated well with previous reports of endothelial damage after phacoemulsification with IOL implantation. The light-adjustable IOL irradiation protocol for adjustment and lock-in procedures did not seem to add to the endothelial damage caused by the cataract surgery, indicating the protocol is safe to the corneal endothelium.
No takes yet. Share an insight, caveat, or question.
Lichtinger et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: