Calcium sulfate doped with dysprosium (CaSO 4 :Dy) has been used for personal dosimetry using the thermoluminescence (TL) technique for decades and its TL properties are well-known. Considering the advantages of the OSL technique (relatively simple and faster readout and detector reusability) and the potential application of CaSO 4 :Dy as an OSL detector, this study aimed to investigate the dosimetric properties of CaSO 4 :Dy using the OSL technique. The present study was carried out using CaSO 4 :Dy pellets prepared at IPEN irradiated and read in a TL/OSL Risø reader with blue light stimulation and Hoya U-340 filter. OSL dosimetric properties essentially dose response, reproducibility, and fading characteristics were evaluated. Moreover, TL and OSL signals were compared to study the correlation between OSL signals and TL peaks. CaSO 4 :Dy OSL response is linear from 0.1 to 10 Gy and shows reproducibility better than 5%. The low-temperature TL peak is unstable; however, the OSL signal seems to be associated with higher-temperature peaks.