Key result
Radiation exposure to anaesthetists over 6 months was 2.32 mSv in the cardiac catheterisation suite and 0.28 mSv for ERCP, both well below the maximum recommended exposure of 20 mSv per year.
Why the study?
What is the level of radiation exposure to anaesthetists during interventional radiological procedures in the ERCP suite and cardiac catheterisation laboratory, and does it exceed safety guidelines?
Observational (n=1,344)
What is the level of radiation exposure to anaesthetists during interventional radiological procedures in the ERCP suite and cardiac catheterisation laboratory, and does it exceed safety guidelines?
Radiation exposure to anaesthetists during interventional radiological procedures, including in the cardiac catheterisation laboratory, remains well below the maximum recommended annual safety limits.
Answers are grounded in this paper and cite their sources.
Reassures low occupational exposure for anaesthetists in these suites; extends single-center data but leaves open multicenter confirmation.
This prospective study determined the level of radiation exposure of anaesthetists during interventional radiological procedures performed in the endoscopic retrograde cholangiopancreatography suite and cardiac catheterisation laboratory and compared it with the current safety guidelines. Anaesthetists wore area-specific lithium fluoride thermo-luminescent dosimeter badges at standardised positions. A total of 1344 procedures were performed over a 6-month period. Anaesthetists were involved in 39/645 (6.0%) procedures associated with ionisation radiation in the endoscopic retrograde cholangiopancreatography suite and 86/699 (12.3%) in the cardiac catheterisation laboratory. The mean (SD) duration of endoscopic retrograde cholangiopancreatography was 54.8 (29.1) min compared with 67.9 (42.8) min for cardiac catheterisation suite procedures (p = 0.058). The mean (SD) fluoroscopy time per procedure for endoscopic retrograde cholangiopancreatography was 5.5 (4.1) min compared with 12 (10.9) min in the cardiac catheterisation suite (p < 0.001). The combined net radiation exposure over 6 months was 0.28 mSv for endoscopic retrograde cholangiopancreatography procedures and 2.32 mSv in the cardiac catheterisation suite. The combined exposure was less than the maximum recommended exposure of 20 mSv per year.
No takes yet. Share an insight, caveat, or question.
Ismail et al. (2009) conducted an observational in Radiation exposure (n=1,344). Cardiac catheterisation procedures vs. Endoscopic retrograde cholangiopancreatography (ERCP) procedures was evaluated on Combined net radiation exposure over 6 months (mSv). Radiation exposure to anaesthetists over 6 months was 2.32 mSv in the cardiac catheterisation suite and 0.28 mSv for ERCP, both well below the maximum recommended exposure of 20 mSv per year.
Synapse has enriched 10 related papers. Consider them for comparative context: