Why the study?
Does a sirolimus-eluting stent with bioresorbable polymer improve long-term clinical outcomes compared to an everolimus-eluting stent with permanent polymer in patients requiring PCI?
Does a sirolimus-eluting stent with bioresorbable polymer improve long-term clinical outcomes compared to an everolimus-eluting stent with permanent polymer in patients requiring PCI?
A sirolimus-eluting stent with a bioresorbable polymer provides comparable long-term safety and efficacy to a permanent polymer everolimus-eluting stent at 5 years.
Supports either BP-SES or PP-EES for PCI; confirms long-term equivalence at 5 years.
AIMS: The aim of this study was to establish the long-term safety and efficacy of a sirolimus-eluting stent with bioresorbable polymer (BP-SES; Ultimaster) by comparison with an everolimus-eluting stent with permanent polymer (PP-EES; XIENCE). METHODS AND RESULTS: CENTURY II (Clinical Evaluation of New Terumo Drug-Eluting Coronary Stent System in the Treatment of Patients with Coronary Artery Disease) is a large-scale, prospective, multicentre, randomised single-blind, controlled, non-inferiority trial conducted at 58 study sites globally, including Europe, Japan and Korea, powered to prove non-inferiority for freedom from target lesion failure (TLF: cardiac death, target vessel-related myocardial infarction [MI] and target lesion revascularisation) at nine months. Patients requiring a percutaneous coronary intervention (PCI) were randomised (1:1) to BP-SES (n=551) or PP-EES (n=550). Freedom from TLF at five years was 90.0% in the BP-SES and 91.1% in the PP-EES group (p=0.54). The patient-oriented composite endpoint (all death, any MI, any revascularisation) was 24.1 and 25.6% (p=0.57) with BP-SES and PP-EES, respectively. The very late stent thrombosis rate from one to five years was especially low at 0.2% in both arms. CONCLUSIONS: This randomised clinical trial showed that the BP-SES stent was non-inferior to the benchmark PP-EES stent for TLF. Safety and efficacy measures were comparable up to five-year follow-up after PCI.
No takes yet. Share an insight, caveat, or question.
Wijns et al. (2018) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: