Why the study?
Does a frame-normalisation correction method prevent overestimation of LVEF and wall motion in patients with marked arrhythmias undergoing equilibrium radionuclide angiography?
Does a frame-normalisation correction method prevent overestimation of LVEF and wall motion in patients with marked arrhythmias undergoing equilibrium radionuclide angiography?
Answers are grounded in this paper and cite their sources.
A simple frame-normalisation method for equilibrium radionuclide angiography prevents the overestimation of LVEF and wall motion in patients with marked arrhythmias.
A simple method of correcting equilibrium radionuclide angiographic data for the effects of arrhythmias is proposed. This involves no further acquisition time or inconvenience to the patient. This correction is necessary in patients with marked arrhythmias as the left ventricular time-activity curve (LVTAC) becomes distorted with greater sampling time in the first few frames of the LVTAC than in later frames. This will show on the LVTAC as greater counts in the first few frames and will lead to an overestimation of left ventricular ejection fraction (LVEF) and wall motion. The method proposed normalises each frame to make the contribution from a non-cardiac region constant over the cardiac cycle. Results of the correction are presented and discussed.
No takes yet. Share an insight, caveat, or question.
Taylor et al. (1979) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: