This is the authors' abstract. We don't add key points for this paper.
Summary Automated haematology analysers determine the volume of red blood cells and provide an erythrocyte volume distribution. For analysis of these distributions, the International Council Tor Standardization in Haematology (ICSH) has recommended methods for fitting a single reference distribution and assessing its goodness of fit (ICSH, 1990). To evaluate the suitability of the ICSH reference method for routine use with haematology analysers, we applied the ICSH method to a reference sample group of 71 healthy individuals and to 618 samples collected from 112 patients. Samples were measured using Sysmex NE-8000 haematology analysers. Each distribution was doubly truncated to eliminate artefactual frequency counts and tested for goodness of fit to a single lognormal distribution (McLaren, Brittenham & Hasselblad, 1986) and mixtures of lognormal distributions (McLaren el al., 1991). Analysis of data from healthy individuals demonstrated the reproducibility of the ICSH reference method. Analysis of data from patients illustrated the ability of the method to detect sequential changes in distributions, providing a useful tool to monitor the effect of therapeutic intervention and blood transfusion.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
McLaren et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: