Key result
Mitral valve annuloplasty using either a complete rigid ring or incomplete flexible band reduced mitral annulus area and resulted in a planar shape compared to pre-surgery.
Why the study?
Does mitral valve repair with annuloplasty alter mitral annulus dynamic morphology in patients with mitral valve prolapse compared to normal subjects, and does the ring type affect these changes?
Cohort (n=64)
Does mitral valve repair with annuloplasty alter mitral annulus dynamic morphology in patients with mitral valve prolapse compared to normal subjects, and does the ring type affect these changes?
Following mitral valve repair for prolapse, the undersizing of the mitral annulus dimensions, rather than the specific type of annuloplasty ring used, is the primary determinant of postoperative annulus function.
Answers are grounded in this paper and cite their sources.
Undersizing drives post-repair annulus changes more than ring type; leaves open whether morphology affects durability in prolapse repair.
AIMS: Mitral valve (MV) repair is the preferred treatment for mitral regurgitation associated with organic MV prolapse (MVP). Our goals were to describe by transthoracic real-time 3D echocardiography (RT3D TTE) the pre-operative changes in mitral annulus (MA) dynamic morphology related to MVP, compared with a normal population, and to evaluate the differential long-term effects induced by annuloplasty, using either an incomplete flexible band or a complete semi-rigid ring. METHODS AND RESULTS: Forty-four patients (62 ± 11 years) with organic MVP and ejection fraction >55% were studied by RT3D TTE the day before MV repair, and 3 and 6 months after (23 patients received a complete rigid ring-CAR, 21 an incomplete flexible band-COS). An age-matched group of 20 normal subjects (57 ± 9 years) was studied as control. After initialization, the MA was tracked frame-by-frame in 3D, and several parameters computed. Differences in MVP vs. controls, vs. pre-surgery, and between rings were tested (P < 0.05). MVP showed enlarged MA resulting in greater area and height during the cardiac cycle, with reduced planarity compared with controls. Annuloplasty resulted in reduced MA area in both CAR and COS, with minimal area change, and planar shape (more evident in CAR than COS). CONCLUSION: The main factor affecting MA function after annuloplasty appears to be the undersizing of the MA dimensions, and not the choice of the ring. This methodology could represent the basis for further evaluation of implanted rings, to provide the surgeon with additional information to be used in the pre-surgical planning and ring selection.
No takes yet. Share an insight, caveat, or question.
Caiani et al. (2011) conducted a cohort in Organic mitral valve prolapse (n=64). Mitral valve repair with annuloplasty (complete rigid ring or incomplete flexible band) vs. Normal subjects and pre-surgery baseline was evaluated on Mitral annulus dynamic morphology (area, height, planarity). Mitral valve annuloplasty using either a complete rigid ring or incomplete flexible band reduced mitral annulus area and resulted in a planar shape compared to pre-surgery.
Synapse has enriched 10 related papers. Consider them for comparative context: