Key result
The ChoRe prototype successfully implanted all 10 artificial chords ex-vivo in bovine hearts with an average procedure time of 3.45 ± 1.44 minutes.
Why the study?
To design and evaluate a new trans-catheter device to place artificial chords in the mitral valve to treat mitral regurgitation caused by chordae tendineae rupture.
Can the ChoRe trans-catheter device prototype successfully attach artificial chordae tendineae to the left ventricular apex and posterior mitral valve leaflet in an ex-vivo bovine heart model?
Can the ChoRe trans-catheter device prototype successfully attach artificial chordae tendineae to the left ventricular apex and posterior mitral valve leaflet in an ex-vivo bovine heart model?
This proof-of-concept ex-vivo study demonstrates the mechanical feasibility of a novel miniature device (ChoRe) for trans-catheter mitral valve chordae tendineae repair, achieving apex and leaflet fixation in an average of 3.45 minutes.
Answers are grounded in this paper and cite their sources.
Ex-vivo feasibility of ChoRe supports device iteration; leaves open in-vivo performance and clinical translation for mitral repair.
This work focuses on the design of a new device (called ChoRe) to place artificial chords in the mitral valve structure during a trans-catheter procedure. The aim of the device is to restore the correct functionality of the valve and solve mitral valve regurgitation, that is, a common consequence of chordae tendineae rupture. An analysis of the requirements was carried out and used to design and develop a first functional prototype. The resulting device was able to connect artificial chords at the posterior leaflet of the mitral valve and at the apex of the left ventricle, also allowing the control of the artificial chord length. The ChoRe was tested ex-vivo in bovine hearts. The qualitative assessment of the ChoRe focused on the performance of the device and preliminary evaluation of the procedure time. Results demonstrated that the device is able to create a top and bottom fixation in an average time of 3.45 ± 1.44 min. Future improvements will focus on enhancing the connection at the leaflet, as well as the overall functionality, in order to guarantee better control of the artificial chord length. This work shows future potentials for more patient-specific treatments in trans-catheter scenarios for mitral valve repair.
No takes yet. Share an insight, caveat, or question.
Culmone et al. (2019) studied Mitral valve regurgitation due to chordae tendineae rupture (ex-vivo bovine heart model) (n=10). ChoRe prototype (3D-printed 2x scaled trans-catheter chordae tendineae repair device) was evaluated on Successful implantation of artificial chord (apex and leaflet fixation) and procedure completion time. The ChoRe prototype successfully implanted all 10 artificial chords ex-vivo in bovine hearts with an average procedure time of 3.45 ± 1.44 minutes.
Synapse has enriched 10 related papers. Consider them for comparative context: