Why the study?
The study evaluated the reliability of the portable Lactate Pro 2 analyzer and its validity compared to a laboratory-based analyzer.
Is the portable Lactate Pro 2 analyzer reliable and valid compared to a laboratory-based analyzer for measuring blood lactate during exercise?
Is the portable Lactate Pro 2 analyzer reliable and valid compared to a laboratory-based analyzer for measuring blood lactate during exercise?
The Lactate Pro 2 analyzer is a reliable and valid tool for identifying training intensities at the lactate threshold, supporting its interchangeable use in practice.
Answers are grounded in this paper and cite their sources.
Portable lactate monitoring may aid field testing in rowing; leaves open broader validity confirmation versus lab standards.
This study evaluated the reliability of the portable Lactate Pro 2 analyzer (LP2), and its validity compared to a laboratory-based analyzer, YSI 1500 Sport (YSI). Blood samples (n = 258) were collected during 44 graded incremental rowing tests, with data from 17 tests used to quantify load, heart rate, and oxygen consumption at lactate threshold. Inter-LP2 reliability was high with coefficient of variation = 3.3%; Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L−1 for blood lactate (BLa) ≤4.0 mmol.L−1 to −1.6 to +1.4 mmol.L−1 for BLa >8.0 mmol.L−1, with minimal systematic bias. Despite LP2 measurements being ~1.42 times higher than YSI measurements, no significant differences (P > .05) were detected when quantifying load, heart rate, or oxygen consumption data at lactate threshold. The results support using different LP2 devices interchangeably, especially when BLa ≤ 8 mmol.L−1, and using the LP2 to identify training intensities at lactate threshold.
No takes yet. Share an insight, caveat, or question.
Crotty et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: