Key result
Biphasic shocks cut defibrillation energy requirements ~77% compared to monophasic shocks.
Why the study?
The electrophysiologic characteristics of fibrillation change with ischemia, and the relative efficacy of biphasic shocks compared to monophasic shocks as fibrillation deteriorates was unclear.
Do biphasic shocks improve defibrillation efficacy compared to monophasic shocks in isolated rabbit hearts after 1-minute ischemic fibrillation?
Do biphasic shocks improve defibrillation efficacy compared to monophasic shocks in isolated rabbit hearts after 1-minute ischemic fibrillation?
Biphasic shocks demonstrate superior defibrillation efficacy over monophasic shocks after 1-minute ischemic fibrillation in a rabbit model, associated with the development of diastolic intervals.
Answers are grounded in this paper and cite their sources.
Biphasic shocks show promise in ischemic VF model; hypothesis-generating, requires confirmation before clinical translation.
Biphasic Shocks and One‐Minute Fibrillation. Introduction: The probability of survival decreases to 70% after 2 minutes of ventricular fibriltation. Bipliasic shocks are more effective than monophasic shocks in terminating short‐duration (<30 sec) ventricular fibrillation. We tested the hypotheses that developing ischemia changes the electrophysiologic characteristics of fibrillation and that the relative efficacy of biphasic shocks increases as electrophysiologic characteristics deteriorate. Methods and Results: Monophasic (12 msec) and biphasic (6/6 msec) shocks (1 to 4 A) were tested in random order in isolated rabbit hearts after 1‐minute ischemic fibrillation. Monophasic action potentials showed only a sporadic occurrence of electrical diastole after 5 seconds of fibrillation (24% of action potentials in the right ventricle and 18% in the left ventricle). After 60 seconds of fibrillation, diastole (17.83 ± 1.14 msec in the right ventricle and 21.52 ± 1.16 msec in the left ventricle) appeared after almost every action potential (P < 0.0001 compared with 5 sec), despite a lack of change in fibrillation cycle length and dominant frequency. Monophasie I50 was 2.89 A, and biphasic I50 was 1.4 A (77% reduction in energy). Normalized curve width decreased 28%. Retrospective analysis showed that shocks delivered early in the fibrillation action potential bad a greater probability of succeeding (89%) than shocks delivered late (30%; P < 0.001). Conclusion: After l‐minute ischemic fibrillation, diastolic intervals occur during fibrillation. Therefore, defibrillation shocks have an approximately 29% probability of interacting with the fibrillation action potential during diastole. At this time, biphasic shocks produced a more deterministic defibrillation threshold and became even more efficacious (I50B/M = 0.48) than at short fibrillation durations (I50 B/M = 0.7).
No takes yet. Share an insight, caveat, or question.
Tovar et al. (2000) studied Ischemic ventricular fibrillation. Biphasic shocks (6/6 msec) vs. Monophasic shocks (12 msec) was evaluated on Defibrillation threshold (I50) (77% reduction in energy). In isolated rabbit hearts after 1-minute ischemic fibrillation, biphasic shocks were more efficacious than monophasic shocks, requiring 77% less energy for defibrillation.
Synapse has enriched 10 related papers. Consider them for comparative context: