A new Time-of-Flight Mass Spectrometer (TOFMS) has been developed that employs an electrospray source and an inductively coupled plasma simultaneously to provide molecular, structural, atomic, and isotopic information from a sample. This paper will focus on characterization of the ICP source and will demonstrate simultaneous operation of the two sources. Preliminary sensitivities and limits of detection for Cr, Co, Ga, As, Ag, Cs, Ho, Hg, Tl, and Bi on the ICP channel are reported. Long-term stability and drift of the ICP channel are also discussed. Simultaneous operation of both channels is demonstrated for cyanocobalamin, myoglobin, and superoxide dismutase. Characterization of the atomic channel of a new dual-source time-of-flight mass spectrometer is described. Focus is given to quantitative capabilities, stability, and isotope precision. Preliminary results are also given for the simultaneous operation of both the atomic and molecular channels.