Key result
The use of ultra-low dose radiation protocols in combination with optimized shielding resulted in an extremely low average operator radiation dose of 0.52 μSv during left atrial ablation procedures.
Why the study?
Does an ultra-low dose radiation protocol with optimized shielding reduce operator radiation exposure during left atrial ablation in patients with atrial fibrillation?
Observational (n=201)
No
Does an ultra-low dose radiation protocol with optimized shielding reduce operator radiation exposure during left atrial ablation in patients with atrial fibrillation?
The use of ultra-low dose radiation protocols combined with optimized shielding results in extremely low scattered radiation reaching the operator during left atrial ablation procedures.
Supports minimal operator exposure in left atrial ablation; leaves open whether protocols alter long-term risks or warrant routine use.
Background Three-dimensional mapping systems and the use of ultra-low dose radiation protocols have supported minimization of radiation dose during left atrial ablation procedures. By using optimal shielding, scattered radiation reaching the operator can be further reduced. This prospective study was designed to determine the remaining operator radiation exposure during left atrial catheter ablations using real-time dosimetry. Methods Radiation dose was recorded using real-time digital dosimetry badges outside the lead apron during 201 consecutive left atrial fibrillation ablation procedures. All procedures were performed using the same X‑ray system (Siemens Healthineers Artis dBc; Siemens Healthcare AG, Erlangen, Germany) programmed with ultra-low dose radiation settings including a low frame rate (two frames per second), maximum copper filtration, and an optimized detector dose. To reduce scattered radiation to the operators, table-suspended lead curtains, ceiling-suspended leaded plastic shields, and radiation-absorbing shields on the patient were positioned in an overlapping configuration. Results The 201 procedures included 139 (69%) pulmonary vein isolations (PVI) (20 cryoballoon ablations, 119 radiofrequency ablations, with 35 cases receiving additional ablation of the cavotricuspid isthmus) and 62 (31%) PVI plus further left atrial substrate ablation. Mean radiation dose measured as dose area product for all procedures was 128.09 ± 187.87 cGy ∙ cm2 with a mean fluoroscopy duration of 9.4 ± 8.7 min. Real-time dosimetry showed very low average operator doses of 0.52 ± 0.10 µSv. A subanalysis of 51 (25%) procedures showed that the radiation burden for the operator was highest during pulmonary vein angiography. Conclusion The use of ultra-low dose radiation protocols in combination with optimized shielding results in extremely low scattered radiation reaching the operator.
No takes yet. Share an insight, caveat, or question.
Schreiber et al. (2021) conducted an observational in Atrial fibrillation (n=201). Ultra-low dose radiation settings and optimized shielding was evaluated on Operator radiation dose measured by real-time dosimetry. The use of ultra-low dose radiation protocols in combination with optimized shielding resulted in an extremely low average operator radiation dose of 0.52 μSv during left atrial ablation procedures.