Key result
The actim Influenza A&B test demonstrated 65% sensitivity and 100% specificity compared with RT-PCR for detecting influenza A and B viruses in respiratory specimens.
Why the study?
Does the actim Influenza A&B test accurately detect influenza A and B compared to RT-PCR in respiratory specimens?
Observational (n=473)
Does the actim Influenza A&B test accurately detect influenza A and B compared to RT-PCR in respiratory specimens?
The actim Influenza A&B test provides reliable positive results due to high specificity, but its low sensitivity necessitates confirmation of negative results with a more sensitive assay like RT-PCR.
Negative results warrant RT-PCR confirmation; leaves open prospective validation across prevalence settings.
Nowadays, influenza antigen detection test kits are used most frequently to detect influenza A or B virus to establish the diagnosis of influenza rapidly and initiate appropriate therapy. This study was conducted to evaluate the performance of the actim Influenza A&B test (Medix Biochemica). Overall, 473 respiratory specimens were analysed in the actim Influenza A&B test and the results were compared with those from an RT-PCR assay; 461 of these samples originated from paediatric patients aged 7 weeks to 6.5 years either with influenza-related symptoms or from the intensive care unit, and 12 samples originated from adults with underlying lung or haematological diseases. Diagnosis of influenza A or B virus could be established using the actim Influenza A&B test (9/473 samples for influenza A virus and 6/473 for influenza B virus). RT-PCR revealed 23 patients with influenza virus (13/473 for influenza A virus and 10/473 for influenza B virus). The sensitivity and specificity of the actim Influenza A&B test were 65 and 100 % compared with the RT-PCR assay. However, 32 external quality assessment samples containing seven different strains of influenza A subtypes H1N1 and H3N2 and the avian H5N1 were detected correctly by the actim Influenza A&B test. No cross-reactivity to a range of bacterial, fungal and other viral pathogens was observed. In conclusion, the actim Influenza A&B test is reliable for positive results due to its high specificity. Nevertheless, negative results from this test need to be confirmed by a more sensitive assay because of the low sensitivity observed with diagnostic samples.
No takes yet. Share an insight, caveat, or question.
Ghebremedhin et al. (2009) conducted an observational in Influenza (n=473). actim Influenza A&B test vs. RT-PCR assay was evaluated on Detection of influenza A or B virus (Sensitivity 65%, Specificity 100%). The actim Influenza A&B test demonstrated 65% sensitivity and 100% specificity compared with RT-PCR for detecting influenza A and B viruses in respiratory specimens.