A simple, precise, and robust RP-HPLC method was successfully developed and validated for the quantitative estimation of ivabradine hydrochloride in commercial tablets.
Answers are grounded in this paper and cite their sources.
May aid ivabradine tablet quality control; leaves open routine lab adoption pending independent validation.
A simple, sensitive, precise and robust reverse-phase high-performance liquid chromatographic method for analysis of ivabradine hydrochloride in pharmaceutical formulations was developed and validated as per ICH guidelines. The separation was performed on SS Wakosil C18AR, 250×4.6 mm, 5 μm column with methanol:25 mM phosphate buffer (60:40 v/v), adjusted to pH 6.5 with orthophosphoric acid, added drop wise, as mobile phase. A well defined chromatographic peak of Ivabradine hydrochloride was exhibited with a retention time of 6.55±0.05 min and tailing factor of 1.14 at the flow rate of 0.8 ml/min and at ambient temperature, when monitored at 285 nm. The linear regression analysis data for calibration plots showed good linear relationship with R=0.9998 in the concentration range of 30-210 μg/ml. The method was validated for precision, recovery and robustness. Intra and Inter-day precision (% relative standard deviation) were always less than 2%. The method showed the mean % recovery of 99.00 and 98.55 % for Ivabrad and Inapure tablets, respectively. The proposed method has been successfully applied to the commercial tablets without any interference of excipients.
No takes yet. Share an insight, caveat, or question.
Srinivasan et al. (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: