Key result
Technological advances like contact force enabled catheters and dedicated electrophysiology labs were strongly associated with lower radiation dose during AF ablation.
Why the study?
AF catheter ablation commonly requires fluoroscopy, but factors associated with lower fluoroscopy doses during cryoballoon or radiofrequency ablation were not fully evaluated.
What factors are associated with lower fluoroscopy doses during atrial fibrillation ablation with cryoballoon or radiofrequency?
Observational
Yes
What factors are associated with lower fluoroscopy doses during atrial fibrillation ablation with cryoballoon or radiofrequency?
Technological advances and a dedicated fluoroscopic environment are strongly associated with lower radiation doses during atrial fibrillation ablation.
Answers are grounded in this paper and cite their sources.
May support prioritizing contact-force catheters and dedicated labs to minimize radiation; leaves open need for randomized confirmation.
AIMS: Atrial fibrillation (AF) catheter ablation is a common procedure requiring in most cases the use of fluoroscopy. We aimed to evaluate the factors associated with a lower dose of fluoroscopy used during AF ablation with cryoballoon or radiofrequency. METHODS: In this prospective European registry, centers were requested to provide procedural characteristics of consecutive AF ablation cases. Lower doses of fluoroscopy were defined as those with dose-area-product (DAP) under the median dose used in the radiofrequency and the cryoballoon ablation groups. RESULTS: in the radiofrequency and cryoballoon groups, respectively. Main factors associated with lower DAP in cryoballoon ablation group were electrophysiology dedicated laboratory (OR 6.04, 95%CI 1.16-31.54; P = .03) and frequent dosimetry report (OR 21.39, 95%CI 5.43-98.54; P = .03). Main factors associated with lower DAP in the radiofrequency ablation group were the use of a chest dosimeter (OR 12.57, 95% CI 2.88-54.90; P = .01), biplane X-ray equipment (OR 3.12, 95%CI 1.89-5.16; P < .01), university hospital (OR 2.10, 95%CI 1.35-3.25; P = .01), electrophysiology dedicated laboratory (OR 2.45, 95%CI 1.48-4.05; P < .01) and use of contact force enabled catheter (OR 22.60, 95%CI 6.82-74.88; P < .01). CONCLUSION: This real-life study of fluoroscopy use during AF ablation provides new data about current practices across European countries. Technological advances and quality of the fluoroscopic environment were the main factors associated with lower radiation dose during AF ablation.
No takes yet. Share an insight, caveat, or question.
Gras et al. (2021) conducted an observational in Atrial fibrillation. Technological advances and quality of fluoroscopic environment (e.g., contact force enabled catheters) vs. Standard environment/equipment was evaluated on Lower dose of fluoroscopy (dose-area-product under the median dose) (OR 22.60, 95% CI 6.82-74.88, p=<0.01). Technological advances like contact force enabled catheters and dedicated electrophysiology labs were strongly associated with lower radiation dose during AF ablation.