Key result
Transapical septal pacing achieves 100% procedural success in canines, but leads dislodge by 12 weeks.
Why the study?
Transvenous lead implantation often faces challenges regarding venous access and lead-related complications, prompting the evaluation of a novel echocardiography-guided percutaneous transapical intraventricular septum pacing approach.
Does echocardiography-guided percutaneous transapical intraventricular septal pacing provide a feasible and safe alternative to transvenous pacing in a canine model?
Does echocardiography-guided percutaneous transapical intraventricular septal pacing provide a feasible and safe alternative to transvenous pacing in a canine model?
Echocardiography-guided transapical intraventricular septal pacing is technically feasible in a canine model, but significant improvements in lead fixation technology are required to prevent long-term dislodgement.
Answers are grounded in this paper and cite their sources.
Acute feasibility shown in canines but dislodgement limits durability; leaves open human translation pending fixation advances.
Background: Transvenous lead implantation often faces challenges regarding venous access and lead-related complications. This study evaluated the feasibility and safety of a novel echocardiography-guided percutaneous transapical intraventricular septum (PTAIVS) pacing approach in an animal model. Methods: Twelve adult dogs underwent PTAIVS lead implantation guided by transthoracic and transoesophageal echocardiography. Leads were introduced percutaneously through the apex and fixed in the mid-to-basal septum. The first six dogs were implanted with a pacing set with a coaxial introducer needle (CareFusion) and a Model 3830 lead (Medtronic), whereas an improved pacing set with an extended helix lead was used in the next six dogs. The first three dogs were used for evaluation of acute gross pathology, with the remaining nine followed up over a period of 4-12 weeks. Pathological examinations were performed at the end of the follow-up period. Results: Successful lead placement without procedural complications was achieved in all dogs, with a mean (± SD) procedural time of 28.8 ± 4.8 min. At implantation, the median (interquartile range) capture threshold was 1.7 (0.85-2.50) V, the R-wave amplitude was 6.80 (6.13-13.00) mV and impedance was 536 (510-922) Ω. In the initial six dogs in which the unmodified pacing system was used, lead dislodgement occurred at the 1-month follow-up. Modifications to the system eliminated dislodgement at 1 month, but varying degrees of displacement were observed by 3 months. Conclusion: This study demonstrates the initial technical feasibility of echocardiography-guided PTAIVS pacing, offering a potential alternative to traditional transvenous methods. However, further refinements are essential to improve long-term lead stability.
No takes yet. Share an insight, caveat, or question.
Qiu et al. (2025) studied Healthy animal model (n=12). Echocardiography-guided percutaneous transapical intraventricular septal (PTAIVS) pacing was evaluated on Procedural success (successful electrode implantation confirmed through unipolar pacing). Echocardiography-guided percutaneous transapical intraventricular septal pacing achieved 100% procedural success in a canine model, though lead dislodgement occurred in all animals by 12 weeks.
Synapse has enriched 10 related papers. Consider them for comparative context: