Key result
Transaortic LV lead placement proves feasible in pigs without significant regurgitation or thromboembolism over six months.
Why the study?
Is direct transaortic placement of a left ventricular lead feasible and safe in a minipig model?
Is direct transaortic placement of a left ventricular lead feasible and safe in a minipig model?
Answers are grounded in this paper and cite their sources.
Direct transaortic placement of a left ventricular lead is feasible in a porcine model, with acceptable lead function and no significant aortic regurgitation or thromboembolism at 6 months.
INTRODUCTION: We tested the feasibility of a new technique of direct left ventricular endocardial lead placement across the aortic valve in a chronic (six month) pig model. The potential for aortic valve damage, systemic embolization, and pacing lead maturation and function within the left ventricle are unknown. METHODS: Ten minipigs were successfully implanted with a transaortic left ventricular lead (Medtronic CapSureFix, Minneapolis, MN, USA) placed in the left ventricular apex via the carotid artery. Each pig received either a polyurethane (n = 5) or silicone (n = 5) lead. Post implant each pig received clopidogrel and aspirin for seven days. After six months all surviving pigs underwent thorough necropsy. RESULTS: Each pig had adequate sensing (12.1 +/- 4 mV) and pacing thresholds (0.79 +/- 0.2 @ 0.5 V) at implant. Postoperatively two pigs died of a respiratory illness. One pig died postoperatively due to sepsis. At the six-month follow-up, all surviving pigs (n = 7) were in a healthy state. Of the pigs without dislodgement (n = 5) there was adequate sensing, but a rise in pacing thresholds. Echocardiography revealed a normal ejection fraction and only trace to mild aortic insufficiency in all pigs. Of the seven surviving pigs there were no thromboembolic events noted. One silicone lead was noted to have thrombosis along the lead screw and shaft. CONCLUSION: Direct transaortic placement of a left ventricular lead is feasible. After six months, there was no significant aortic regurgitation and no evidence of thromboembolism despite no anticoagulation. Lead function was acceptable and only one silicone lead (and no polyurethane lead) was noted to have significant thrombosis.
No takes yet. Share an insight, caveat, or question.
Reinig et al. (2007) studied Left ventricular endocardial pacing (animal model) (n=10). Transaortic left ventricular lead placement vs. Polyurethane vs silicone lead was evaluated on Feasibility, aortic valve damage, systemic embolization, and pacing lead maturation and function. Direct transaortic placement of a left ventricular lead was feasible in a pig model, with no significant aortic regurgitation or thromboembolism after six months despite no anticoagulation.