Key result
High ultrasound image gain (70 dB) overestimated carotid intima-media thickness compared to the default 50 dB setup and reduced measurement precision.
Why the study?
The study aimed to investigate how ultrasound image settings affect carotid intima-media thickness and diameter estimation performed with Carotid Studio.
How do different ultrasound image settings affect the estimation and repeatability of carotid IMT, diameter, and distension using semi-automatic B-mode image processing?
Cross-Sectional (n=14)
No
How do different ultrasound image settings affect the estimation and repeatability of carotid IMT, diameter, and distension using semi-automatic B-mode image processing?
Ultrasound settings, particularly high gain and persistence filters, significantly affect the precision and reproducibility of semi-automated carotid biomarker assessments, highlighting the need for standardized imaging protocols.
Answers are grounded in this paper and cite their sources.
High gain may overestimate carotid IMT in practice; leaves open optimal settings for standardized semi-automated protocols.
Introduction This study aims to investigate how image settings affect carotid Intima-Media-Thickness (IMT) and diameter estimation performed with Carotid Studio (Quipu-Srl). Methods Fourteen healthy volunteers (24.4 ± 4.4 years; eight men) were recruited. Longitudinal scans 3-cm-deep of common-carotid-arteries with the same region-of-interest were obtained by expert operator, starting from a “default-setup” (central Time Gain Compensation-TGC; gain: 50 dB; no persistence filter) and adjusting one parameter at a time (oblique TGC; gain: 10 dB increments from 30 to 70 dB; persistence filter). The acquisition was performed twice, and repeatability was expressed for each setup as Coefficient of Variation (CV). Significant changes (p ≤ 0.05) compared with the “default-setup” were analysed by Wilcoxon signed-rank-test. Results Sixty and 70 dB gain resulted in overestimating IMT (from 0.460 ± 0.001 to 0.506 ± 0.030 mm and 0.538 ± 0.011 mm, p = 0.002 and 0.023 respectively vs. default-50 dB), underestimating the mean diameter (from 6.044 ± 0.040 mm to 5.763 ± 0.004 mm, p = 0.008 and to 5.698 ± 0.225 mm, p = 0.002), and for 70 dB, increasing IMT CV from 2.14 ± 1.92% to 6.85 ± 5.23% (p = 0.01) and diameter CV from 1.79 ± 1.22% to 4.77 ± 4.71% (p = 0.05). Distension CV increased from 5.71 ± 4.31% to 10.84 ± 5.91% (p = 0.041) and to 12.59 ± 7.97% (p = 0.021) for 60 and 70 dB respectively. Persistence and TGC did not affect repeatability, whereas diameter (5.955 ± 0.033 mm, p = 0.005) and distension (from 0.816 ± 0.019 mm to 0.655 ± 0.018 mm, p = 0.002) values were underestimated when persistence filter was active. Discussion Precision and reproducibility of carotid parameters, automatically assessed by Carotid Studio, are affected by high gain levels, due to image grey-level saturation. Persistence filter activation leads to underestimate diameter and distension, possibly due to mathematical operator behaviour and image temporal filtering. Since ultrasound setup significantly affects carotid parameter assessment, it should be reported and replicated in follow-up scans.
No takes yet. Share an insight, caveat, or question.
Francesconi et al. (2019) conducted a cross-sectional in Healthy volunteers (n=14). High ultrasound image gain (70 dB) vs. Default setup (50 dB gain) was evaluated on Carotid Intima-Media Thickness (IMT) (p=0.023). High ultrasound image gain (70 dB) overestimated carotid intima-media thickness compared to the default 50 dB setup and reduced measurement precision.