Centrifugal pumps are superior to roller pumps for extended support durations in terms of pump-induced haemolysis. In this study, we evaluated the commonly used Biomedicus BP 50 and compared it with the Jostra Rotaflow and a standard roller pump in an in vitro test circuit. Each circuit was run for a six-day period and repeated five times. Plasma haemoglobin values showed the roller pump to become more haemolytic than the Biomedicus (p = 0.022) and the Rotaflow. A statistically significant difference between the Biomedicus and the Rotaflow was observed on day six of the trial (p = 0.016), with the Rotaflow showing lower levels of haemolysis than the Biomedicus. These results support the use of the new generation centrifugal pump, the Rotaflow, as a suitable device for short-term ventricular assist.
No takes yet. Share an insight, caveat, or question.
Bennett et al. (2004) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: