The detection of carbon monoxide (CO) is of paramount importance for environmental monitoring, industrial safety, and public health. This study presents an all-fiber gas concentration monitoring technique based on tunable diode laser absorption spectroscopy (TDLAS), offering low gas consumption, high time resolution, high stability, and high precision. A 1-m-long negative curvature anti-resonant hollow core fiber (HCF) with a core diameter of110~μm is used for both gas containment and optical transmission. Experimental and theoretical simulations confirm that the response time of the system reaches 1.27 s at an overpressure of 98 kPa. Furthermore, the system achieves a relative standard deviation (RSD) of less than 2.5% and a minimum detection limit (MDL) of 0.220 ppm under optimal overpressure conditions. In addition, a spatial-resolved scanning imaging is demonstrated for the CO concentration distribution of 190-μm leakage point, enabling clear identification of the topographical features. This technology has potential in the fields of environmental monitoring, industrial safety, and public health.