Key result
Changes in right ventricular apical pacing impedance correlated with changes in left ventricular ejection fraction (r = 0.45, P = 0.002) and NYHA class in heart failure patients.
Why the study?
Do serial changes in right ventricular pacing lead impedance predict changes in LVEF and NYHA functional class in heart failure patients?
Observational (n=67)
Do serial changes in right ventricular pacing lead impedance predict changes in LVEF and NYHA functional class in heart failure patients?
A decrease in bipolar pacing impedance at the right ventricular apex, but not the outflow tract, is associated with worsening LVEF and NYHA class in heart failure patients.
Answers are grounded in this paper and cite their sources.
RV impedance trends may aid HF surveillance in paced patients; leaves open utility pending prospective validation.
Pacing impedance has been proposed to monitor the clinical status of patients with congestive heart failure (CHF). This study examined whether changes in right ventricular (RV) pacing impedance correlate with changes in left ventricular ejection fraction (LVEF) and New York Heart Association (NYHA) functional class during long-term follow-up in pacemaker recipients with CHF. The study included 67 patients, 70 +/- 12 years of age, in NYHA class II or III, and with a mean LVEF = 29 +/- 8% at implant. LVEF, NYHA class, and bipolar pacing impedance at the RV outflow tract (RVOT) and apex (RVA) were measured at implant and at 3, 6, 9, and 12 months of follow-up. At implant, impedance was similar in RVOT (548 +/- 115 Omega) and RVA (571 +/- 174 Omega). Between implant and 3 months, mean impedance decreased (P < 0.0001) at both the RVOT (472 +/- 62 Omega) and RVA (488 +/- 86 Omega), LVEF increased (43 +/- 14%, P < 0.0001), and the NYHA class decreased from 2.4 +/- 0.5 to 2.1 +/- 0.6 (P = 0.0001). Changes in RVA impedance correlated with changes in LVEF (r = 0.45, P = 0.002). A 50 Omega decrease in RVA impedance corresponded to a 3% decrease in LVEF. RVA impedance decreased significantly as NYHA class increased from I to IV (P = 0.04). There was no correlation between impedance measured at the RVOT and LVEF or NYHA class. A decrease in bipolar pacing impedance at the RVA was associated with worsening LVEF and the NYHA class. The use of pacing impedance to monitor the clinical status in CHF is dependent on the RV pacing site.
No takes yet. Share an insight, caveat, or question.
Stambler et al. (2005) conducted an observational in Congestive heart failure (CHF) (n=67). Right ventricular apical (RVA) pacing lead impedance was evaluated on Correlation between changes in RVA impedance and changes in LVEF (r = 0.45, p=0.002). Changes in right ventricular apical pacing impedance correlated with changes in left ventricular ejection fraction (r = 0.45, P = 0.002) and NYHA class in heart failure patients.
Synapse has enriched 10 related papers. Consider them for comparative context: