Key result
Intrauterine transvenous cardiac pacing at 200 beats/min decreased right ventricular output from 107 to 78.3 mL/kg per min without increasing tricuspid valve regurgitation.
Why the study?
Does intrauterine transvenous cardiac pacing affect right ventricular output in fetal lambs?
Does intrauterine transvenous cardiac pacing affect right ventricular output in fetal lambs?
Intrauterine transvenous cardiac pacing is feasible in a fetal lamb model but decreases right ventricular output without increasing tricuspid regurgitation.
Answers are grounded in this paper and cite their sources.
Alerts to hemodynamic risks with fetal pacing; hypothesis-generating in lambs and leaves open human translation.
To evaluate the feasibility of intrauterine transvenous cardiac pacing, the right ventricular output was measured during pacing in six fetal lambs. Under maternal anesthesia, the uterus was opened, and, under local anesthesia, the pacing lead (Medtronic Capsure SP4023) was inserted via the fetal left internal jugular vein. Right ventricular output was estimated using an Aloka SSD-730 ultrasound device, and tricuspid valve regurgitation was evaluated with an Aloka SSD-880 using the transuterine approach. The ultrasonic right ventricular cardiac output was measured under three different conditions: (1) with the tip of the pacing lead in the superior vena cava (control); (2) with the tip of the pacing lead in the right ventricle; and (3) with pacing at 200 beats/min. The right ventricular output decreased when the pacing lead was inserted into the right ventricle, as well as during pacing at 200 beats/min ([1] = 107 +/- 13.2 mL/kg per min; [2] = 73.8 +/- 17.5 mL/kg per min; and [3] = 78.3 +/- 23.6 mL/kg per min). Tricuspid regurgitation did not change under any of the conditions tested. Intrauterine transvenous cardiac pacing was successfully achieved. Insertion of the pacing lead into the right ventricle decreased the ventricular output without increasing tricuspid valve regurgitation.
No takes yet. Share an insight, caveat, or question.
Kikuchi et al. (1995) studied Fetal complete heart block (n=6). Intrauterine transvenous cardiac pacing vs. Pacing lead in superior vena cava (control) was evaluated on Right ventricular output. Intrauterine transvenous cardiac pacing at 200 beats/min decreased right ventricular output from 107 to 78.3 mL/kg per min without increasing tricuspid valve regurgitation.