Cryogenic microcalorimeters can be made sensitive enough to measure the energy deposited by a single particle or X-ray photon with an accuracy of about one electron volt. It may also be possible to construct detectors of several kilograms mass whose resolution is only a few times worse than this. Data from relatively crude test devices are in good agreement with thermal performance calculations, and a total system noise of 11 eV FWHM has been obtained for a silicon detector operating at 98 mK. We have observed 35 eV FWHM for 6 keV X-rays with a different device.