Why the study?
Does adjusting to unipolar ventricular pacing prevent inappropriate ICD shocks due to double counting in a patient with a separate pacemaker and ICD?
Does adjusting to unipolar ventricular pacing prevent inappropriate ICD shocks due to double counting in a patient with a separate pacemaker and ICD?
In patients with separate pacemakers and ICDs, adjusting to unipolar ventricular pacing can resolve inappropriate ICD shocks caused by double counting of bipolar pacing spikes.
Answers are grounded in this paper and cite their sources.
May avert inappropriate shocks from device interaction in select cases; leaves open generalizability beyond this report.
This report describes a 20-year-old man with complex congenital heart disease and inappropriate epicardial implantable cardioverter defibrillator (ICD) shocks secondary to double counting of ventricular-paced spikes and QRS complexes from a separate bipolar epicardial dual-chamber pacemaker. Adjusting to a unipolar paced-ventricular mode resolved any double counting via pacemaker-ICD interaction.